User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Fluke 80F-15:High Voltage Divider DATE: 17-Feb-95 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 5 NUMBER OF LINES: 60 CONFIGURATION: Fluke 5700A CONFIGURATION: Datron 1281 STANDARD: ESI SR1 10Mê ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R N P U M F W 1.002 ASK+ X 1.003 IEEE [@1281]*RST 1.004 STD ESI SR1 10Mê 1.005 HEAD ÍÍ INITIAL CONDITIONS ÍÍ 1.006 DISP This procedure requires the use of the ESI 10Mê 1.006 DISP resistor, DTB# 18-59 1.007 HEAD {} 1.008 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.009 HEAD {º Divider Calibration º} 1.010 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.011 HEAD ÍÍ Divider Calibration ÍÍ 1.012 JMP 2.001 1.013 EVAL 2.001 DISP Connect the ESI standard resistor to the Datron DMM 2.001 DISP HI and LO INPUTS. 2.002 DISP Connect the divider box from the 80F-15 to the 2.002 DISP terminals of the Standard Resistor. 2.003 DISP Connect the probe tip from the UUT to the 2.003 DISP 5700A HI OUTPUT. 2.004 DISP Connect the ground clip from the UUT to the 2.004 DISP LO OUTPUT from the 5700A. 2.005 IEEE [@1281]DCV AUTO 2.006 5700 100V S 2W 2.007 ACC 1000 0.01% 2.008 IEEE [@1281][D5000]RDG?[I] 2.009 MATH MEM=100/MEM 2.010 MEME 2.011 MEMC I 0.070% 100ÿVolts 3.001 5700 500V S 2W 3.002 ACC 1000 0.01% 3.003 IEEE [@1281][D5000]RDG?[I] 3.004 MATH MEM=500/MEM 3.005 MEME 3.006 MEMC I 0.070% 500ÿVolts 4.001 5700 1kV S 2W 4.002 ACC 1000 0.01% 4.003 IEEE [@1281][D5000]RDG?[I] 4.004 MATH MEM=1000/MEM 4.005 MEME 4.006 MEMC I 0.070% 1000ÿVolts 5.001 HEAD {} 5.002 DISP {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 5.002 DISP {º THIS COMPLETES THE VERIFICATION º} 5.002 DISP {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 5.003 END